Futrex Body Fat and fitness Analyzer -5500A/WL User Manual

Page 35

Advertising
background image

35

PROBLEM

PROBABLE CAUSE

CORRECTIVE MEASURES

3) Bad Repeatability

a) Inconsistent Measurement Technique

b) Incorrect Zero Adjustment Procedure

c) Not measuring at the proper location
on the arm

a) Ensure that there us enough pressure

applied between the Light Wand and the

subject’s arm

b) When taking a measurement, ensure

that the Silver Stripe on the Light Shield is

pointing towards the subject’s shoulder

(the Silver Stripe on the Light Wand

should be aligned with the Silver Stripe on

the Light Shield).

c) Always use the Light Shield to block

external light.

d) Avoid taking measurements under

strong lights (e.g. direct sunlight) as errors

may occur

e) When Zero Adjusting, ensure that the

silver line on the Light Wand is aligned

with the silver line on the Optical

Standard.

f) When Zero Adjusting, make sure that

the Light Shield has been removed from

the Light Wand. This is to ensure that the

Light Wand is properly seated in the

Optical Standard.

g) Always use the Biceps Locator to find

the measurement site. Instructions can

be found on page 12.