Electrical Safety Analyzer

Controlling the Analyzer Remotely

55

After performing and saving all the tests for one Test
Record ID, enter a new Test Record ID to close the
previous record and start a new one.

Note

A previously closed record can not have
measurements or ECG signals added to it. Data
can only be stored under the most recently
opened record.

Viewing Memory Data

Previously stored data for any Test Record is recalled to
the screen through the setup menus. To recall data:
1. Press

Q

.

2. Press the softkey labeled More to reveal additional

menu selections.

3. Press the softkey labeled View Memory.
4. Select the desired Test Record by pressing G or H

to scroll through the records listed in the display.

5. Press the softkey labeled View to view the data

stored for the selected record.

If the stored data is more than can fit on one screen, press
the softkey labeled Next Page to view the additional data.

Deleting Data from Memory

To delete a Test Record and its associated data from
memory:

1. Press

Q

.

2. Press the softkey labeled More to reveal additional

menu selections.

3. Press the softkey labeled View Memory.
4. Select the desired Test Record by pressing G or H

to scroll through the records listed in the display.

5. Press the softkey labeled Delete.

Note

All Test Records can be deleted at this point by
pressing the softkey labeled Delete All.

6. When

Delete? appears in the display, press the

softkey labeled Delete to remove the record from
memory. Otherwise press the softkey labeled Cancel
to return to the Test Record list and leave the
selected record in memory.

Note

Deleting the last or current record does not open
the previous record for additional data storage.
Only a new Test Record can be opened for data
recording after deleting the last record.

Controlling the Analyzer Remotely

Fluke Biomedical Ansur test automation software allows a
solutions-based approach to complete testing of the
medical device under test (DUT). Ansur helps create
standard work using the test template/sequence (which is