Operation

Operation Instructions

2

2-7

Figure 2-6. The Model 35360A TRACKER Display Front Panel

Power

The POWER button toggles the system power on or off.

Test/Detector

The TEST/DETECTOR button displays and sequences the Test/Detector screens.

Units Select

The UNITS SELECT button displays the Units Select screen. Subsequent presses toggle the
measurement units between dose and dose rate units.

Air Density

The AIR DENSITY button displays the Air Density screen. Subsequent presses advance the cursor
through the temperature, pressure, and Air Density Correction Factor (ADCF) fields.

Gain Select

The GAIN SELECT button displays the Gain Select screen. Holding the GAIN SELECT key for three
seconds toggles the 35360A TRACKER Display between the Gain Setting and Normal Operation
modes. In the Gain Setting Mode, pressing the GAIN SELECT button advances the cursor through the
Front Panel Gain Factor (FPGF) fields.

Measure Select

The MEASURE SELECT button displays the Measurement screen. Subsequent presses toggle the
display mode between absolute and percentage presentations.

Reset

The RESET button zeroes all measurement readings when dose units are selected. This button
displays the Measurement screen if it is not already displayed.

Up

The UP button function depends on the displayed screen. It increments the temperature, pressure, or a
front panel gain. It toggles the state of the air density correction annunciator or the front panel gain
annunciator. For the Test/Detector screen or the Measurement screen, it displays the previously
displayed screen.

Down

The DOWN button function depends on the displayed screen. It decrements the temperature,
pressure, or a front panel gain. It toggles the state of the air density correction annunciator or the front
panel gain annunciator. For the Test/Detector screen or the Measurement screen, it displays the next
screen.

TURN POWER OFF

PRIOR TO CONNECTING

AND DISCONNECTING

DETECTOR CABLE

TRACKER DISPLAY

Model 35360A