Theory of Operation

Basic Operating Quantities

3

3-7

3.6 Basic Operating Quantities

Table 3-3. Basic Operation Quantities

Quantity Description

Units

Comments

I

Current nC/Min

Electrical current delivered to readout in

response to radiation incident upon

detector.

Q

Charge nC

Electrical charge delivered to readout in

response to radiation incident upon

detector.

ADC

A-D Counts

Counts

Electrometer output level returned by

A-D Converter during dose or dose rate

measurement.


ADZ


A-D Zero


Counts

Electrometer output level returned by

A-D Converter while operating in the

dose rate measurement mode and in

the absence of radiation. A-D Zero

levels are field programmable via

Customization Software.


QG


Charge Gain


nC / Count

Overall electrical charge gain of the

electrometer / A-D combination when

measuring in the dose mode. Charge

gain values are field programmable via

Customization Software.


IG


Current Gain


(nC/Min) / Count

Overall electrical current gain of the

electrometer and A-D combination

when measuring in the dose rate mode.

Current gain values are field

programmable via Customization

Software.

ICCF

Ion Chamber

Calibration

Factor

Ion Chamber

Calibration Units

/ nC

Specific R/nC values are supplied by

factory for each of the five ion chamber

channels. Values are field

programmable via Customization

Software.

DUCF

Display Units

Conversion

Factor

Display Units

/ Radiation Calibration

Units

Factory default of 0.879 rad/R is

supplied. Value is field programmable

via Customization Software.

ADCF

Air Density

Correction

Factor

None

Calculated by instrument and depends

on front panel Temperature and

Pressure settings.

FPGF

Front Panel

Gain Factor

None

User specified via front panel gain

setting.