RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 28

 PRINCIPLE

RPM4 A70M/A20M-AF is designed to support the calibration and test of a very wide variety of
test ranges in absolute and gauge measurement modes.

The AutoRange function simplifies the task of selecting the Q-RPT best suited to cover a
specific test range and setting measurement parameters appropriately for that specific range.
These are set automatically based on operator entry of desired measurement mode, pressure unit
of measure and range full scale. The selections and settings made by AutoRange are summarized
in Table 4.

A range set up using AutoRange and all its range specific settings, may be saved for recall
using [SETUP], <1range> (see Section 3.4.1). Saved ranges are available for later selection
using [RANGE] (see Section 3.3.1). If a range created by AutoRange is not saved, it is
overwritten by the next AutoRange or erased when another range is selected.

Press [ENT] from the main run screen in Rate DISPLAY mode to use the [AutoTest]

function to AutoRange RPM4 based on DUT tolerance (see Section 3.3.10).

Table 4. Settings made by AutoRange

SETTING/

SELECTION

PURPOSE

SETTING BY

[AUTORANGE]

SETTING BY
[AUTOTEST[

(see Section 3.3.10)

RELATED

SECTION

Unit

Set pressure unit of
measure

Operator specified
during AutoRange

Operator specified in
AutoTest definition

3.3.2

Mode

Set pressure
measurement mode
(absolute, gauge)

Operator specified
during AutoRange

Operator specified in
AutoTest definition

3.3.3

Full scale

Establish span
maximum

Operator specified
during AutoRange

Maximum pressure of
AutoTest definition

None

Q-RPT

Select most
appropriate Q-RPT to
cover range and mode.

Q-PRT with lowest
default full scale that is
greater than the
AutoRange full scale

Q-PRT with lowest
default full scale that is
greater than the
maximum pressure of
AutoTest definition

1.2.2.1

Upper limit

Set maximum pressure
alarm limit.

105 % of AutoRange
full scale pressure

105% of maximum
pressure of AutoTest
definition

3.4.4

Resolution

Set display resolution

10 ppm of AutoRange
span or
1 ppm of default Q-
RPT span, whichever is
larger

10% of DUT tolerance
or
1 ppm of default Q-
RPT span, whichever is
larger.

3.4.2

Stability Limit

Set Ready/Not Ready
stability criterion.

50 ppm of AutoRange
span or 2 ppm of
default Q-RPT span,
whichever is larger

10% of DUT tolerance
or
2 ppm of default Q-
RPT span, whichever is
larger

3.4.3,
3.2.1