4. REMOTE OPERATION

Page 89

© 2007 DH Instruments, a Fluke Company

PRR
Purpose

Read the next available Ready condition, pressure measurement, rate and on-board
barometer reading.

Query “PRR?”
Classic “PRR”
Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT

Remarks The

next

available

Ready condition, Q-RPT pressure measurement, rate of pressure change,

and barometric pressure is replied in the current pressure unit of measure. . This can take up
to the current read rate period (1.2 seconds by default). Each data field is separated by a
comma, and is returned in the following order:

ready, pressure UNITS, rate UNITS/s, atm UNITS

Here are the field descriptions:

ready:

Three character ready status field. Refer to the “SR” command.

pressure:

The measured pressure for the active Q-RPT in the current pressure unit.
This is followed by the current pressure unit.

rate:

The measured rate of pressure change for the active Q-RPT in the current
pressure unit per second. This is followed by the current pressure unit of measure.

atm:

The pressure measured by the RPM4 on-board barometer in the current
pressure unit (and always absolute). This is followed by the current pressure
unit. Not all RPM4s are equipped with an on-board barometer. This field is
missing if the RPM4 is not equipped with an on-board barometer.

Example
(enhanced)

Query sent:
Query reply:

“PRR?”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPa a”
“R,2306.265 kPaa,0.011 kPa/s”

(no barometer)

Example
(classic)

Query sent:
Reply:

“PRR”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPa a”

See Also

“PR”, “QPRR”, “SR”
3.1.1, 3.2.1, 3.5.7.1

QPRR
Purpose

Read the last Q-RPT pressure measurement, pressure rate and on-board barometer output
immediately.

Query “QPRR?”
Classic “QPRR”
Optional Suffix

“n”

The active Q-RPT is assumed if no suffix is given.
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT

Remarks

The last measured Ready/Not Ready condition, active Q-RPT pressure measurement, rate of
pressure change, and barometric pressure is replied in the current pressure unit of measure
immediately. This program message is useful when a rapid response of measured pressure
is needed. It does not result in quicker pressure measurements. Each data field is separated
by a comma, and is returned in the following order:

Ready, pressure UNITS, rate UNITS/s, atm UNITS

Here are the field descriptions:

ready:

Three character ready status field. Refer to the “SR” command.

pressure:

The measured pressure for the selected Q-RPT in the current pressure unit.
This is followed by the current pressure unit and measurement mode.

rate:

The measured rate of pressure change for the active Q-RPT in the current
unit of pressure per second. This is followed by the current pressure unit.

atm:

The pressure measured by the RPM4 on-board barometer in the current
pressure unit (but always absolute). This is followed by the current pressure
unit. Not all RPM4s are equipped with an on-board barometer. This field is
missing if the RPM4 is not equipped with a barometer.

Example
(enhanced)

Query sent:
Query reply:

“QPRR?”
“R,2306.265 kPa a,0.011 kPa/s,97.000 kPa a”
“R,2306.265 kPa a,0.011 kPa/s” (no barometer)

Example
(classic)

Query sent:
Reply:

“QPRR”
“R,2306.265 kPa a,0.011 kPa/s,97.000 kPa a”

See Also

“PR”, “PRR”, “SR”