RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL

© 2007 DH Instruments, a Fluke Company

Page 66

customizes the [UNIT] function key to support his/her most commonly used units. Up to six
units can be included in the [UNIT] screen.

 OPERATION

To customize the [UNIT] function key, press [SPECIAL], <6PresU> from the MAIN RUN
screen. The display is:

1.

Entry field to specify which unit position (1 to 6) of the [UNIT]
function key menu is to be changed.

Set up unit #6

Enter the number of the unit position that you would
like to change. The display becomes:

Unit#6 1SI 2other

3altitude 4user

Select the desired pressure unit category (SI units include units based on SI units such as
mmHg), then select the desired unit from the unit menu. Note: <3altitude> is not available in
this RPM4 model.

The units of measure available are listed in Table 7.

Table 7. UNIT function - available units of measure

<1SI>

<2OTHER>

<3ALTITUDE

>

<4USER>

<1Pa>

<2hPa>
<3kPa>

<4MPa>

<5mbar>

<6bar>

<7mmHg>

<8mmWa>

<1psi>
<2psf>

<3inHg>

<4inWa>

<5kcm2>

Not available

in this RPM4

model

If <4user> is selected, the user unit must be defined. The display is:

1. Entry

field.

Define user unit:

1.000000 unit/Pa

Enter the number of user units per Pascal (Pa) in the entry field. Pressing [ENT] defines the
user unit and returns to the <Set up unit #n> screen.

The user defined unit label can be customized to any alphanumeric, four character label

using the remote command “UDU” (see Section 4.4.4).

See Section 7.1.1 for the pressure unit conversion factors used by RPM4.

3.5.7

<7INTERNAL>

 PURPOSE

To view, set, adjust, and maintain various aspects of RPM4’s internal operation.

 OPERATION

To access the internal selections press [SPECIAL],
<7internal>. The display is:

1baro 2ReadRt 3RPT2x

4lo vnt 5log