Fluke 43B User Manual

Page 47

Advertising
background image

Specifications

Current Probe 80i-500S

3

43

Current Probe 80i-500S

Safety Characteristics

Designed for measurements on 600 V rms Category III. Pollution degree 2,
double or reinforced insulation requirements in accordance with:

EN/IEC

61010-2-032

ANSI/ISA

S82

CSA-C22.2

No.1010.1-92

UL1244

Electrical Specifications

Current range ........................................................................ 1A to 500 A rms
AC current over range limit ..............................................................700 A rms

Maximum 10 minutes, followed by removal from current carrying
conductor for 30 minutes.

Output Signal............................................................................ 1mV AC/A AC

Accuracy

5 to 10 Hz

1 to 500 A .............................................................................-3 dB typically

10 to 20 Hz

1 to 300 A ..........................................................................................

±

5 %

300 to 400 A ....................................................................................

±

15 %

400 to 500 A ....................................................................................

±

25 %

20 to 45 Hz

1 to 500 A ..........................................................................................

±

5 %

45 to 65 Hz

1 to 20 A ...............................................................

±

5 % of reading + 0.3 A

20 to 100 A ........................................................................

±

5 % of reading

±

3

°

phase shift

100 to 500 A ......................................................................

±

2 % of reading

±

5

°

phase shift

65 Hz to 3 kHz

1 to 50 A ..............................................................................

±

(5 % + 0.4 A)

50 to 500 A ........................................................................................

±

5 %

Influence of temperature on accuracy ................... <0.15 % per 10

°

C (18

°

F)

Altitude

During operation .................................................................2.0 km (6560 feet)
While stored .....................................................................12 km (40 000 feet)

Advertising