Gentek 3940302 User Manual

Page 24

Advertising
background image

19

Chapter 3: Getting Started

Filter Type

Appropriate Use

HEPA

2' 3707900
3' 3707901
4' 3707902

HEPA filters are high-efficiency particulate
air filters having a particulate removable
efficiency of 99.99% for particles with a
diameter of 0.3 micron.

Organic Vapor

Activated Carbon

2' 3937200
3' 3937300
4' 3937400

Adsorbs organic compounds designated by
NIOSH guidelines as acceptable for use with
chemical cartridge-type respirators.
Concentrations in the enclosure’s work area
must not exceed the IDLH for the chemical
and the exhaust from the enclosure must not
exceed the TWA.

Formaldehyde

(Formalin)

2' 3937201
3' 3937301
4' 3937401

Formaldehyde requires the use of an
impregnated carbon for the treatment of
formaldehyde. Concentrations in the
enclosure’s work area must not exceed the
IDLH for the chemical and the exhaust from
the enclosure must not exceed the TWA.

Ammonia and

Amines

2' 3937202
3' 3937302
4' 3937402

Treats ammonia, low molecular weight
amines and other bases designated by NIOSH
as acceptable for use with ammonia cartridge
type respirators. Requires the use of an
impregnated carbon. Concentrations in the
enclosure’s work area must not exceed the
IDLH for the chemical and the exhaust from
the enclosure must not exceed the TWA.

Important: Carbon filters do not provide any
particulate protection, but provide odor control.

Connecting the Electrical Supply
Source to the Filtered Enclosure

115V Models

Simply connect the 115V power cord supplied to the IEC electrical
supply plug on the back of the enclosure. If using at 50 Hz
operations, blower performance maximum airflow will be reduced
by 17%.

For the remote blower used for XPert Stations, follow the wiring
diagram in Appendix C and wire the blower wires in the electrical
junction box located behind the front panel. These wires are
switched by the electrical relay via the blower switch.

HEPA

Carbon

Carbon

Carbon

XPert

System

XPert

Station

Advertising