Chapter 6: Maintaining Your Filtered Enclosure

Measure of Line Voltage and Current

Smoke Test to determine proper airflow patterns

Lighting Intensity Test (when appropriate)

Noise Level Test (when appropriate)

Vibration Test (when appropriate)

Fluorescent Light Replacement

1. Disconnect the power.
2. Locate the small light reflector located under the control

panel shown in Figure 6-2 and Figure 6-3.

3. Remove the light reflector support by removing two

Phillips screws on the bottom of the light reflector.

4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp and light reflector in

reverse order.

6. Power the unit up and try the new fluorescent lamp.

Motorized Impeller Replacement

The motorized impeller must be replaced as a complete unit.
When the motorized impeller is replaced, the capacitor may also be
replaced. See Appendix A for Replacement Parts Diagram. See
Figure 6-5 for an isometric view of the motorized impeller plenum
assembly. The HEPA filter rests on top of the motorized impeller
assembly.

1. Wear appropriate personal protective equipment to

decontaminate the filtered enclosure and then unplug from the
electrical outlet.

2. Remove HEPA filter per the HEPA filter removal procedure

outlined in this chapter.

3. Consult the wiring diagram in Appendix C of the manual and

disconnect all the wires of the motorized impeller. Be sure to
connect wires on the new motor in the same way the old motor
was wired.

4. Remove four screws in the motor bracket that hold motor

bracket to the supports with the vibration isolation mounts.
Remove the motor and bracket.

WARNING: High-speed blower. Never operate
impeller with housing off.

5. Replace the capacitor with a new one of equal voltage and

capacity.

!

XPERT

FILTERED
BALANCE

SYSTEM