Gentek 3940302 User Manual
Page 59
54
Chapter 6: Maintaining Your Filtered Enclosure
To start the calibration, use the Labconco 1000 Manual and enter
the “CALIBRATIION” mode on the display. Follow the
instructions on the display and alter the low and high exhaust
volumes with the speed control. The following suggested inflow
face velocity speeds are recommended to successfully calibrate.
Typical low air alarms are set 10-30 fpm below operational speeds.
Typical yellow cautions are set 3-10 fpm above the red low air
alarm. Be sure to enter “Done” after completing the calibration in
“SETUP.” Once completed, go to “RUN” and hit “ENTER.”
Low Air Alarm
Set Point
Nominal Enclosure
Operating In Flow
Speed
Low
Calibration
Set Point
High
Calibration
Set Point
50 fpm
60 fpm
50-60 fpm
100-120 fpm
60-70 fpm
80 fpm
50-70 fpm
100-140 fpm
70-90 fpm
100 fpm
60-80 fpm
110-140 fpm
Factory Settings
60 ± 10 fpm
90 ± 10 fpm
60 ± 10 fpm
90 ± 10 fpm