REPLACEMENT OF THE PREFILTER, POSTFILTER, AND RO MEMBRANE ELEMENTS

17

Rev B

6. Remove exhausted prefilter and discard.

7. If Changing Membrane Elements:

A. Remove membrane sumps. Remove and discard used

elements.

B. Remove new elements from packaging.

C. Lubricate element O-rings, brine seals, and sump O-rings

with silicone lubricant.

D. Securely insert O-ring end of elements into manifold. See

Figure 8.

E. Replace sumps and hand-tighten.

8. Sanitize the system.

A. The manifold should be positioned flat with the sump

connections facing up.

Figure 8

CAUTION:

The person handling the filters and membrane elements

must have clean hands to keep the system sanitized. The use of
sterile/latex gloves is recommended.

CAUTION:

Elements contain a foodgrade preservative. The use of

sterile/latex gloves is highly recommended.

NOTE:

The system should be sanitized whenever a membrane

element or filter is replaced.

Prefilter Center Opening

Membrane Element Locations