Chapter 2: Prerequisites

Enclosure

Width Model

Description

Face

Velocity

(fpm)

Exhaust

Volume

(CFM)

Noise

Pressure

db(A)

Max.

Equivalent

Resistance
of 6" Duct

(Ft)

Max.

External

Static

Pressure

2'

2' Purifier Class I Enclosure
2' Purifier HEPA Filtered
Enclosure

60
75
80
90

100
105

85

110
115
130
145
155

48-53
49-55
50-56
51-57
53-58
54-59

400
220
200
140

80
70

.20"
.19"
.18"
.16"
.13"
.12"

3'

3' Purifier Class I Enclosure
3' Purifier HEPA Filtered
Enclosure

60
75
80
90

100
105

130
165
175
200
220
230

48-53
52-56
53-57
55-58
58-61
60-64

130

65
50
35
25
20

.15"
.12"
.11"
.09"
.08"
.07"

4'

4' Purifier Class I Enclosure
4' Purifier HEPA Filtered
Enclosure

60
75
80
90

100
105

175
220
235
265
295
310

49-54
54-58
58-61
60-63
62-66
64-67

100

60
50
35
25
20

.20"
.19"
.18"
.16"
.13"
.12"


Proper blower selection can be determined from these exhaust
requirements and the total system static pressure loss. For outside
exhaust, the enclosure must be connected to either a dedicated
blower or a house exhaust system.

Labconco offers accessory remote blowers listed in Chapter 7.
Contact Labconco for blower sizing assistance.

If the enclosure is connected directly to a house exhaust system, an
adjustable damper (or valve) must be installed to control the
airflow properly. This is equally important when a house exhaust
system is controlling multiple filtered enclosures. See Chapter 7
for accessory adjustable damper ordering information.

Electrical Requirements

Standard duplex electrical receptacles should be nearby for
connecting the filtered enclosure, or other equipment. The
enclosures include iris pass-throughs to allow electrical cords
through the back of the enclosure without leaving a large hole for
contaminants to escape.

Space Requirements

The dimensions for the different models are shown in Appendix B:
Dimensions

!