Chapter 4: High Performance Features and Safety

21. Lamp Ballast (Not Shown) for the fluorescent lamp is

located behind the control panel.

22. UV (Ultraviolet) Lamp (Not Shown), found only on

Purifier Class I Enclosure, allows the operator to surface
disinfect the work area of the enclosure when it is not in
use. The UV will not penetrate the plane of the sash and
lower airfoil.

23. Accessory Work Surface (Not Shown) is dished and

contoured to fit the dimensions of the enclosures to contain
spills.

24. Accessory Odor-Control Carbon Filters are available for

nuisance odor control. Accessory carbon filters are
designed to remove trace amounts of noxious fumes;
capacity is not adequate to remove continuous fume
generation. The carbon filters are supported above the
HEPA filter and the gasket side is down and compressed by
the weight of the carbon filter. Carbon filters are listed in
Chapter 7. See Figure 4-2.

25. Accessory Exhaust Connection Kits are available to duct

the Purifier Class I Enclosure and Purifier HEPA Filtered
Enclosure to the outside. The exhaust connection allows
the operator additional protection from hazardous fumes or
vapors and should be used if the carbon filter capacity is
not adequate. Each kit includes an epoxy-coated steel
exhaust transition adapter with manual adjustable damper
designed for 6" diameter PVC duct. See Figure 3-3 in
Chapter 3.

26. Upper Diffuser Screen provides protection for the HEPA

Filter and (LED) airflow sensor. See Figure 4-1 and 4-2.

27. HEPA Filter Access Port to access HEPA filter during

filter changing procedures. For harmful powders and
particulates, the true bag-in/bag-out procedure is used to
properly dispose of the contaminated HEPA filter. See
Figure 4-2.

28. Utility Shelves (Not Shown) allow the user to hold utensils

and printers on shelves inside the enclosure. See Chapter 7.