Maintenance 49, Periodic checks 49, Maintenance – Miele G 7823 User Manual

Page 49: Periodic checks

Advertising
background image

Periodic checks

This washer-disinfector should be
inspected in accordance with local and
national safety regulations after every
1000 operating hours, or every 6
months by a Miele approved service
technician.

This maintenance will cover the
following:

– Electrical safety tests

– Door mechanism and door seal

– Any screw connections and

connectors in the wash cabinet

– Water inlet and drainage

– Internal and external dispensing

systems

– Spray arms

– Filter combination

– Sump including drain pump and

non-return valve

– Steam condenser

– All mobile units, modules and inserts

and where applicable:

– The drying unit

– Any printer connected to the

machine

The following operational tests will be
carried out within the framework of the
maintenance:

– A programme will be run as a test

run

– Thermo electrical measurements will

be taken

– Seals will be tested for water

tightness

– All relevant measuring systems will

be safety tested including error
message displays.

Maintenance