Invacare 211EX User Manual

Page 17

Advertising
background image

17

PART NO. 1093845

VENTURE™

OPERATION

3. If using an oxygen cylinder, check that the regulator is

properly attached before opening the cylinder valve.

4. Connect the Venture DODD to the regulator by performing

one (1) of the following:
A. If using an Oxygen Cylinder and Regulator:

On a 20 psig regulator, connect the tubing (supplied with

the Venture DODD) from the barb fitting of the regulator

to the barb inlet fitting on the Venture DODD.
On a 50 psig regulator, connect the threaded (DISS) fitting

(supplied with the Venture DODD) to the regulator DISS fitting.

NOTE: The 50 psig system should ONLY be used with the threaded

(DISS) fitting and not the 12-inch connecting tube, as the pressure may

be too great for the tube to stay connected.

B. If using a Liquid Oxygen Reservoir:

Directly connect the Venture DODD to the oxygen

source using a DISS fitting.

5. Connect the nasal cannula tubing to the barbed oxygen outlet

fitting on the Venture DODD and adjust cannula for comfort.

WARNING

When using nasal or masked type administering

equipment, oxygen or air tubing MUST be routed and

secured properly to ensure that the tubing does NOT

become entangled, kinked or severed.

NOTE: Use ONLY a NASAL CANNULA with PULSE MODE

delivery, NOT a mask. A mask may NOT fit your face well enough for the

unit to sense your inhalation.
NOTE: A special nasal cannula is NOT required. Any standard flow

crush-proof nasal cannula can be used.
NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.
NOTE: The cannula tubing can be up to 50 feet long to function properly

in the pulse mode. The continuous flow mode is NOT affected by the

length of the canula tubing.