Troubleshooting – Invacare 211EX User Manual

Page 25

Advertising
background image

25

PART NO. 1093845

VENTURE™

TROUBLESHOOTING

PROBLEM

POSSIBLE CAUSE

SOLUTIONS

External input will not

power device (Cont.).

"Low" battery light

comes on too soon after

battery replacement.

External power adapter

plug does not go into

jack in unit.

Unit pulses at times

other than inhalation.

Continuous flow out of

cannula when in pulse

mode.

Check that power adapter

used within specification for

operating voltage of external

power supply. Call your

Home Health Care Dealer

or Service Representative

IMMEDIATELY.
Check battery

manufacturer's

expiration date. See

Prolonging the Life of the

Battery in the MAINTE-

NANCE Section of this

manual. Call your Home

Health Care Dealer or

Service Representative

IMMEDIATELY.
Call your Home Health

Care Dealer or Service

Representative

IMMEDIATELY.
Refer to SPECIFICATION

Section for external

power adapter connector

size.
Check if YELLOW and

GREEN O

2

LEDS illumi-

nate. If so, see indicator

description in this section.

Call your Home Health

Care Dealer or Service

Representative

IMMEDIATELY.
Verify compatibility of

model and pressure.

Call your Home Health

Care Dealer or Service

Representative

IMMEDIATELY.

Incorrect power adapter.

Battery life expired.

Jack or plug damaged.

Non-compatible adapter

plug.

Unit may be in backup

Pulse Mode.

Operating pressure of

oxygen supply source too

high.

NOTE: If any of the above problems occur and the solutions do not work,

please contact your Invacare Home Health Care Dealer for service.