Troubleshooting – Invacare 211EX User Manual

Page 23

Advertising
background image

23

PART NO. 1093845

VENTURE™

TROUBLESHOOTING

This is normal

operation.

Turn unit OFF, check

cylinder pressure is above

refill limit (500 psig) and

post valve is open, check

regulator is set to correct

flowrate. If problem still

exists, call your Home

Health Care Dealer or

Service Representative

IMMEDIATELY and go to

continuous mode if

oxygen supply is adequate.

Breathe normally. Check

cannula for kinking and

check tubing connection.

Breathe normally. Check

cannula for kinking and

check tubing connection.

Turn unit off then on

again. If problem still

exists, call your Home

Health Care Dealer or

Service Representative

IMMEDIATELY and go to

continuous mode if

oxygen supply is adequate.

Replace battery as soon

as possible.

Replace battery or turn

unit OFF to enable

continuous mode if

oxygen supply is adequate.

O

2

LED - (GREEN)

Blinking.

PROBLEM

POSSIBLE CAUSE

SOLUTIONS

Indicates Oxygen Pulse Flow

in normal pulse delivery.

Indicates Venture DODD

not sensing presence of

oxygen in pulse delivery.

Indicates oxygen pulse

delivery, but Venture DODD

not sensing inhalation.

Indicates no pulse

delivery, Venture DODD

not sensing inhalation.

Possible internal problem.

End user possibly not

receiving oxygen.

Battery low.

Approximately two (2) hours

of battery use time left.

Approximately 30 minutes

of battery use time left.

O

2

and LED -

(GREEN and YELLOW)

Blinking

LED - (YELLOW)

Blinking and

Intermittent Beep

LED - (YELLOW)

Solid and Continuous

Beep

O

2

LED - (GREEN)

LED - (RED)

Blinking, Three (3) Beeps

at Beginning.

LED - (RED) Solid,

Six (6) Beeps at Beginning.

Advertising