TROUBLE

REMEDY

PROBABLE CAUSE

CHILLER DOES NOT

OPERATE, BUT COOLING

LIGHT IS “ON”AND SAFETY

LIGHT IS “ON”.

A. Unit is operating under high

pressure conditions.

A. Check for dirty condenser fins or

obstruction of chiller air intake.

Press high pressure manual reset

switch.

B. Unit is operating under low

pressure conditions.

B. Check sight glass, if bubbles are

seen, leak check unit and

recharge with refrigerant.

or

If no bubbles are seen, replace

defective TXV.

C. Unit is operating under low

level conditions.

C. Add liquid to reservoir via fill port.

or

If level is above switch, replace

float switch.

D. Unit is operating under low

temperature conditions.

D. Low or no process liquid flow.

Ensure that there is adequate flow

through system plumbing.

or

Process liquid is too cold, below

35_F. Increase thermostat

setting.

or

Defective thermostat, replace.

When servicing a REMCOR Chiller, it is important to note all information provided on the DATA PLATE located

in the upper rear of the unit. If technical assistance is needed, the REMCOR Service Technician will need this

information along with any description of the problem(s) you are encountering. The serial number and other

information will also be required when ordering replacement parts and any other Warranty Claims.