General information, Introduction, Unpacking and inspection – Cornelius CH Series User Manual

Page 3: Design data, Chiller installation

Advertising
background image

1

92266

GENERAL INFORMATION

INTRODUCTION

The REMCOR ”CH”Series Recirculating Liquid Chiller is designed to provide an accurate, reliable, and user--

friendly system for cooling a continuous flow of pure liquid and keep that liquid at a constant temperature in vari-

ous closed loop or tank cooling applications.

The ”CH”Series Chiller consists of an air--cooled refrigeration system housed in a sturdy sheet metal frame and

cabinet. A standard pump and insulated liquid reservoir package provides a complete liquid cooling and circu-

lating system.

The ”CH”Series Chiller is designed to operate in a clean laboratory or industrial environment where ambient

temperatures range from 40 to100°F (5 to 38°C). With proper installation, operation, and maintenance, the

”CH”Series Chiller will provide years of trouble free service.

UNPACKING AND INSPECTION

This unit was thoroughly inspected before leaving the factory and the carrier has accepted and signed for it.

Any damage or irregularities should be noted at the time of delivery and immediately reported to the carrier. Re-

quest a written inspection report from the Claims Inspector to substantiate any necessary claims. In the event

that an immediate replacement is necessary, please contact REMCOR Chiller Sales at 1--800--551--4423.

DESIGN DATA

Table 1. Design Data

CH951

CH1500

CH1502

CH1503

Cooling Capacity:

BTU/hr (W) at 80_F(27_C) and

70_F (21_ C) Liquid Temperature.

12,000 (3,515)

18,000 (5,272)

18,000 (5,272)

18,000 (5,272)

Compressor Horsepower

1 (.746 kW)

1 1/2 (1.12 kW) 1 1/2 (1.12 kW) 1 1/2 (1.12 kW)

Electrical Data:

Voltage/Phase/Hertz/Amperage

230/1/60

11.0 Amps

230/1/60

13.0 Amps

230/3/60

16.4 Amps

460/3/60

10.0 Amps

Refrigerant Type:

R134a

R22

R22

R22

Reservoir Capacity

6.0 gallons (22.7

liters)

6.0 gallons (22.7

liters)

6.0 gallons (22.7

liters)

6.0 gallons (22.7

liters)

Physical Dimensions, Width ¢ Depth
¢ Height:

22I¢ 26.5XI¢

38.25I

(56cm ¢ 67cm

¢ 97cm)

22I¢ 26.5XI¢

38.25I

(56cm ¢ 67cm

¢ 97cm)

22I¢ 26.5XI¢

38.25I

(56cm ¢ 67cm

¢ 97cm)

22I¢ 26.5XI¢

38.25I

(56cm ¢ 67cm

¢ 97cm)

Fittings:

Process Connections

3/4” FPT (S/S)

3/4IFPT (S/S)

3/4IFPT (S/S)

3/4IFPT (S/S)

CHILLER INSTALLATION

Location of Chiller

THE CHILLER MUST BE LOCATED NEAR A PROPERLY GROUNDED ELECTRICAL OUTLET. THE CIR-

CUIT SHOULD BE FUSED AND NO OTHER ELECTRICAL APPLIANCE SHOULD BE CONNECTED TO THE

CIRCUIT. ALL ELECTRICAL WIRING MUST CONFORM TO NATIONAL AND LOCAL ELECTRICAL

CODES.

The Chiller must be located in a well ventilated, indoor area where ambient temperatures will remain above

40_F (5_C) and will never increase above 100_F (38_C). To obtain optimum cooling capacity, the ambient

temperature should be at or below 80_F (27_C).