SECTION 7—TROUBLESHOOTING

Get-U-Up

Lift

24

Part No 1148115

SECTION 7—TROUBLESHOOTING

NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.

SYMPTOMS

FAULTS

SOLUTION

Stand up lift feels loose.

Mast/base joint loose.
Tie-rods are loose.

Tighten the bolt, washer and locknut that secure the
mast to the base.
Refer to Assembling the Patient Lift on page 10.

Casters/brakes noisy or stiff.

Fluff or debris in bearings.

Refer to Hydraulic Pump on page 25.

Noisy or dry sound from pivots.

Needs lubrication.

Refer to Hydraulic Pump on page 25.

Lift arms will not lower in uppermost position.

Lift arms require a minimum weight
load to lower from the uppermost
position.

Pull down slightly on the lift arms.