Test 1: +7 dbm input level, Hp 3400b rms voltmeter or equivalent – Symmetricom 58502A User Manual

Page 38

Advertising
background image

Chapter 2 Performance Tests

58502A Performance Tests

2-12

User’s Guide

Test 1: +7 dBm Input Level

1

Apply power to the 58502A and observe successful completion of
self-test.

2

Set the signal generator to:
FREQUENCY: 10 MHz
LEVEL: +7 dBm

3

Connect the signal generator output to the 58502A input A or B rear-
panel input connector as shown in Figure 2-6.

4

Connect the RMS voltmeter via the 50

load to the 58502A Output 1

as shown in Figure 2-6.

Figure 2-6. Input Level Response Test Setup

5

Set the RMS voltmeter to 3V Full Scale Range.

6

Starting with Output 1, verify that the RMS voltage on each of the
twelve Outputs is from 0.707 to 1.12 Vrms.

7

Record the reading of each output channel on the Performance Tests
portion of the 58502A Performance Test Record in the Test 1 section,
located on page 2-14.

HP 3400B
RMS Voltmeter
or equivalent

50

Ω Feedthrough

(HP 10100C)

BNC Cable

BNC
Cable

HP 8663A
Synthesized Signal Generator
or equivalent

Output

Outputs

1 - 12

N-to-BNC

Connector

(1250-0780)

Input

A or B

A

B

1

2

3

4

5

6

7

8

9

10

11

12

5 VRMS MAX

100-240 VAC 50/60GHz

40 VA Max

!

!

!

58502A
Distribution
Amplifier