Chapter 2 Performance Tests

58502A Calibration (Option 010 Only)

2-18

User’s Guide

Figure 2-7. 58502A Calibration Setup

4

Observe Counter reading.

If difference from 10 MHz is

<

5

×

10

8

(0.5 Hz), then perform the

“Software Oscillator Adjustment” procedure below.

Software Oscillator Adjustment

1

Use the terminal and serial port to send the following SCPI command
to ascertain the EFC value currently applied:

DIAG:CAL:ROSC:EFC:ABS?

HP 53131A
Universal
Counter
or equivalent

BNC Cable

BNC Cable

Outputs

1 - 12

HP 24542U

Interface

cable

or equivalent

RS232C

Port

To any

available

Output

A

B

1

2

3

4

5

6

7

8

9

10

11

12

5 VRMS MAX

100-240 VAC 50/60GHz

40 VA Max

!

!

!

58502A
Distribution
Amplifier

Terminal or PC

Ref. In
50

Ω Feedthrough

(HP 10100C)

5071A

PRIMARY FREQUENCY STANDARD

Continuous
Operation

HP 5071A
Primary Frequency
Standard or equivalent

Channel 1

Port1

10 MHz External

Reference Out