Test 2: +15 dBm Input Level

1

Using the same test setup (see Figure 2-6) and the RMS voltmeter still
set to 3V Full Scale Range, adjust the signal generator to:

LEVEL: +15 dBm

2

Verify that the RMS voltage on each of the twelve Outputs is from
0.707 to 1.12 Vrms.

3

Record the reading of each output channel on the Performance Tests
portion of the 58502A Performance Test Record in the Test 2 section,
located on page 2-15.

This completes the performance tests.