Temperature setting, Chamber temperature, Key lock function – Sanyo MDF-U32V User Manual

Page 19

Advertising
background image

17

TEMPERATURE SETTING

Chamber temperature

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the

sequence indicated in the table. The example in the table is based on the assumption that the desired

temperature is -75

o

C.

Note: The unit is set at the factory that the chamber temperature -80

o

C.

Table 1. Basic operation sequence (Example: Chamber temperature -75

o

C)

Description of operation

Key operated

Indication after operation

1

Turn the power switch ON.

----

The current chamber temperature is
displayed.

2 Press SET key.

SET

The second digit is flashed.

When pressed, the figure of settable
digit changes.

3

Set to -75 with the numerical value
shift key and digit shift key.

When pressed, the settable digit
is shifted.

4 Press SET key.

SET

Set temperature is memorized and the
current chamber temperature is
displayed.

Note:

• The temperature set mode returns to the temperature display mode automatically when 90 seconds has

passed without any key operation.

• Although the value of the chamber temperature setting can range from -50

o

C to -90

o

C, the guaranteed

temperature when there is no load is -85

o

C when the ambient temperature is 30

o

C.

Key lock function

This unit is provided with the key lock function. When the key lock is ON, change of temperature setting

through the key pad is not available. The key lock is set in OFF at the factory.

Display Mode

Function

L 0

Key lock is OFF

Enable to change of temperature setting

L 1

Key lock is ON

Disable to change of temperature setting

Table 2. Procedure for key lock setting (change from key lock OFF to key lock ON)

Description of operation

Key operated

Indication after operation

1

----

The current chamber temperature is
displayed.

2 Press digit shift key for 5 seconds.


The first digit is flashed.

3

Press numerical value shift key
and scroll the figure to 1.

When pressed, the figure of settable
digit changes.

4 Press SET key.

SET

The key lock is set to ON.
The current chamber temperature is
displayed.



Advertising