Change of compressor delay time – Sanyo MDF-U32V User Manual

Page 25

Advertising
background image

23

CHANGE OF COMPRESSOR DELAY TIME

The delay time of high and low stage side compressor can be changed to reduce the load on the power

line and to facilitate the start-up (reset) of the freezer after power failure.

The example in the table is based on the assumption that the delay time is changed to 4 minutes.

(The delay time is set in 2 minutes at the factory.)

Note:

• The delay time should be same for high stage side and low stage side compressors.

• The setting range for delay time is between 2 and 15 minutes. The cool down of chamber temperature

may be slow when the setting of delay time is over 5 minutes, depending on the installation environment.

There is no need changing the delay time when the capacity of power source is enough.

Table 8. Changing procedure for delay time (change from 2 minutes to 4 minutes)

Description of operation

Key operated

Indication after operation

1

----

The current chamber temperature is
displayed.

2

Press numerical value shift key for

5 seconds.

The first digit is flashed.

3

Set the figure to F05 with the
numerical value shift key.

When pressed, the figure of settable

digit changes.

4 Press SET key.

SET

The current delay time is displayed.

The first digit is flashed.

5

Set the figure to 004 with the

numerical value shift key.

When pressed, the figure of the first
digit changes.

6 Press SET key.

SET

The delay time is memorized and
the current chamber temperature is
displayed.

• The compressor starts to operate with the delay time set by the above procedure at the time of power

on or after power failure. For the low stage side compressor, however, its start up may be behind the

delay time because the compressor is affected by the chamber temperature and temperature of cascade

condenser installed in the freezer.