MONITOR OF FREEZER STATUS

The freezer has a function to monitor the running status of the unit as shown in table 7 below. This is to

notice the running status getting worse (not failure).

1. The STATUS lamp is lit when one of the running status shown in table 7 is detected.

2. The S code (--X, X: 1 to 3) is displayed on the temperature indicator by pressing STATUS key when

the STATUS lamp is lit.

3. Pressing the STATUS key again causes current chamber temperature display on the temperature

indicator. (The indication returns to the chamber temperature display automatically when no key is

operated for 90 seconds.)

Table 7. Monitor of running status

Kind of

function

Status Indication

If this status

continues

Remedy

Notice of

abnormal

ambient

temperature

When the ambient

temp. is over about

35

o

C or lower than

about 0

o

C.

STATUS lamp

lights.

“--1” is displayed.

Decrease of cooling

performance or

durability of

refrigerating circuit.

Recheck of

air-conditioning of

installed site.

Notice of low

voltage

When the power

source voltage is

less than about 97V

when the rated

voltage is 115V or

less than about 195V

when the rated

voltage is between

220 and 240V.

STATUS lamp

lights.

“--2” is displayed.

Abnormal heat at

power supply outlet

or degrade of

starting performance

of refrigerating

circuit

Use of dedicated

power source.

Notice of

overload

condition

When the running

rate of refrigerating

Circuit is more than

usual.

STATUS lamp

lights.

“--3” is displayed.

Decrease of cooling

performance or

durability of

refrigerating circuit.

1. This is likely

happened when

large amount of

materials are stored.

2. Check of ambient

temp., voltage, and

sealing of

outer/inner door.

Note:

• The S code displayed on the temperature indicator is changed every a few seconds if two or three

status shown in the above table are detected at the same time.

--1

--2

--3

--1 repeated

• The monitoring function has no buzzer operation and no safety operation.

• In the case of multiple indication of S code, follow the remedy for each status.