3. OPERATING INSTRUCTIONS page 15

Model BM-286A

Low

Alarm:

Low alarm level in CPS, is the point at which the unit will generate a FAULT

condition. Enter a value between 0 and 1,000. The counts from each detector are

compared to this value; if the count from any detector falls below this value the

unit will enter a FAULT mode.

High

Alarm:

High alarm level in CPS, is the point at which the unit will generate a FAULT

condition. Enter a value between 500 and 32,000. The counts from each detector

are compared to this value; if the count from any detector exceeds this value the

unit will enter a FAULT mode.

Variance

Time:

This menu parameter sets time per variance test, in seconds. Enter a value

between 10 and 100. Refer to Variance (section 3.5.1.) for details.

Number of Trials:

This menu parameter sets the number of iterations for the variance test. Enter a

value between 3 and 10.

Units of Weight:

This menu parameter selects either pounds or kilograms as the unit of weight for

display and report purposes. Minimum weight is always in pounds, and specific

activity is always in pCi/gram regardless of this selection.

Reports:

This menu parameter enables writing a report to disk. The choice is from three

options: ALL, ALARM, or NONE. Refer to Reports (section 3.6.) for details.