5. MAINTENANCE page 36

Model BM-286A

5.3. DISCRIMINATOR ADJUSTMENT

The Upper and Lower Level Discriminators (ULD & LLD) set the maximum and minimum

energy levels at which the system accepts counts. The LLD must be set high enough to prevent

system noise from being amplified and counted.

The DOS machine sets the levels on the RCS-120 discriminator card but gets its values from the

NT machine.

The upper and lower level discriminators’ adjustment range is 0 keV to approximately 1.5 MeV.

These levels are controlled by entering the Utilities Menu and pressing the Detectors, Scale and

Disc. button.

The Set Discriminator box and detector counts will appear with the previous upper and lower

settings. Using the up and down arrows the values for upper and lower can be changed but are

not sent to the DOS machine until the send button is pressed. Observe the counts in the detector

boxes to insure you have not reached the lower level noise threshold.

The approximate energy level in keV can be obtained by multiplying the settings by 150. The

exact value may vary slightly due to variations in the detectors and electronics. For initial set-up,

a good starting point is:

NOTE:

If either level is changed you must always

recalculate all waste stream efficiencies.

The unit will also acquire a new background when leaving the Utilities Menu.

Always run a variance after adjusting the levels.

SETTINGS FOR PLASTIC DETECTORS

LLD

=

0.3 volts

=

45 keV

ULD

=

10.0 turns

=

1.5 MeV