3. OPERATING INSTRUCTIONS page 19

Model BM-286A

Alarm

Type:

Alarm type selects the method of generating a radiation alarm.

The available options are:

1. NSIG: ((N*1background) + background)

2. DPM: Disintegrations per minute above background

3. SPACT: Specific Activity in pCi/gram

Alarm

Message:

This parameter entry is the message that will be displayed if the alarm level is

exceeded. This may be up to 22 alpha-numeric characters. This message will be

included on the data sample information and in the reports.

Clean

Message:

This parameter entry is the message that will be displayed if the alarm level is not

exceeded. This may be up to 22 alpha-numeric characters. This message will be

included on the data sample information and in the reports.

Maximum

Weight:

Enter a maximum allowable weight, in pounds. This is used to determine if the

item to be scanned is within the waste stream weight limits.