Dental, Head dental – Siemens syngo CT 2005C User Manual

Page 86

Advertising
background image

86

Head

Dental

This is the scan protocol for the syngo Dental CT appli-
cation package. It is used for evaluation and reformat-
ting of the upper and lower jaws.

It enables the display and measurement of the bone
structures of the upper and lower jaw as the basis for
planning in oral surgery.

Indications:

Spiral mode for dental studies.

A range of 5 cm will be covered in 27 sec.

Load the study into the application “syngo Dental CT”.
For further information, please refer to the chapter
"syngo Dental CT".

Dental

kV

130

Effective mAs

45

Rotation time

1.0 sec

Slice collimation

1.0 mm

Slice width

1.25 mm

Feed/Rotation

2.0 mm

Pitch Factor

1.0

Increment

0.5 mm

Kernel

H60s

CTDI

Vol

10.25 mGy

Effective dose

Male: 0.13 mSv

Female: 0.16 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 86 Friday, April 8, 2005 9:55 AM