System Sensor DH400ACDCP User Manual

Page 4

Advertising
background image

[5.4.2] Installation For Ducts More Than 8 Feet Wide
NOTE:
To install inlet tubes in ducts more than 8 feet wide,
work must be performed inside the air duct. Sampling of air
in ducts wider than 8 feet is accomplished by using the ST-10
inlet sampling tube. If the tube is shorter than the width of
the air duct, install the end plug into the inlet tube as shown
in Figure 4 and support the end opposite the duct detector.
Install the inlet tube as follows:
1. Drill a

3

/

4

˝ hole in the duct directly opposite the hole

already drilled for the inlet tube.

2. Slide the inlet tube with the flange into the centermost

housing bushing. Position the tube so that the arrows
point into the air flow. Secure the tube flange to the
housing bushing with two #6 self-tapping screws.

3. From inside the duct, couple the other section of the

inlet tube to the section already installed using the

1

/

2

˝

conduit fitting supplied. Make sure that the holes on
both of the air inlet tubes are lined up and facing into
the air flow.

Figure 4. Air duct detector inlet sampling tube:

INLET

TUBE

END

PLUG

AIR HOLES

ARROWS

MUST FACE

INTO AIR FLOW

AIR FLOW DIRECTION

FLANGE

H0108-00

AIR FLOW

DIRECTION

DETECTOR

HOUSING

DOTS INDICATE POSITION OF

SAMPLING TUBE HOLES

AIR FLOW

DIRECTION

DETECTOR

HOUSING

INLET

TUBE

EXHAUST

TUBE

AIR FLOW

DIRECTION

EXHAUST

TUBE

INLET

TUBE

INLET

TUBE

EXHAUST

TUBE

DETECTOR

HOUSING

AIR FLOW

DIRECTION

EXHAUST

TUBE

INLET

TUBE

DETECTOR

HOUSING

A.

B.

C.

D.

HORIZONTAL MOUNTING OF HOUSING

VERTICAL MOUNTING OF HOUSING

Figure 5. Tube mounting configurations with varying air flow direction:

H0109-00

4. Trim the end of the tube protruding through the duct so

that 1˝ to 2˝ of the tube extend outside the duct. Plug
this end with the end plug and tape closed any holes in
the protruding section of the tube. Be sure to seal the
duct when the tube protrudes.

NOTE: An alternate method to using the ST-10 is to use two
ST-5 inlet tubes. Remove the flange from one of the tubes
and install as described above. After the installation, use
electrical tape to close off some of the sampling holes so
that there is a total of 10 to 12 holes spaced as evenly as
possible across the width of the duct.
Air currents inside the duct may cause excessive vibra-
tion, especially when the longer sampling tubes are used.
In these cases a 3˝ floor flange (available at most plumb-
ing supply stores) may be used to fasten the sampling tube
to the other side of the duct. When using the flange/con-
nector mounting technique, drill a 1˝ to 1

1

/

4

˝ hole where

the flange will be used.

D400-13-00

4

I56-0555-010R

Advertising