Model 4501-XX Toxic Gas Sensor Module

Page 15 of 35

6.5 CALIBRATION SUB-MENU










































* 4501-03 –

Oxygen Calibration: If area is known to be safe with adequate air flow, you may use the surrounding air as the calibration gas.
If Oxygen levels are unknown, it is recommended to use a Zero Air gas standard.

When calibrating an Oxygen module, there is no zero gas process in the calibration procedure. Only span adjustments can be
made. Normal ambient Air = 20.9% Oxygen.

Begin calibration procedure using the magnet on the Mode key until CALIB is displayed.
Enter to begin procedure. Select cal gas value (20.9), Enter to confirm.
Apply span gas for 3 minutes or use surrounding air if acceptable.
Select Enter when done, and procedure is complete.

NOTE

The Analog output values to be transmitted during the calibration procedure are selected in

the maintenance configuration. Normal 20.9% value provides 17.3 mA.

Function

Display

Description

Reference

M

E

 

M

E

 

M

E

 

M

E

 

M

E

 

M E

M E

M

E

 

M

E

 

M

E

 

M

E

 

H2S

C 25PPM

Operation: Apply calbration gas.

H2S

CAL-FAIL

Operation: No calibration gas applied, or sensor did not respond
correctly.

(Any)

Operation: Hold magnet over Mode Switch for ten senconds to
abort calibration

Sub C

H2S

WAIT-300

Operation: Five minute time out before sensor is returned to
service.

Enter

H2S

C 25PPM

Banner: Apply 25PPM gas, then enter <E> to calibrate gas
sensor

Sub B

H2S

CXXXPPM

Operation: As gas is applied the reading will increase - wait 3
minutes

Enter

H2S

CAL-OK

Calibration Passed - now remove gas

Mode

H2S

XXXPPM

Default Display

Banner: Apply zero gas, enter <E> when done

Mode

H2S

CALIB:--

Key

Banner: Calibration Aborted

H2S

WAIT-300

Operation: Five minute time out before sensor is returned to
service.

Down

H2S

Adj-SPAN

Operation: User adjustable value

Enter

H2S

25PPM-SP

Operation: Ready for user adjustment

Enter

H2S

10PPM-SP

Banner: Select span, enter <E> when done

Down

H2S

25PPM-SP

Operation: Change Span Gas Value to 25PPM

Sub A

Sub A

Sub B

H2S

CXXX PPM

Operation: Apply calbration gas.

Enter

Mode Function - Calibrate

Enter

H2S

Cal Zero

Banner: Select span, enter <E> when done*

Enter

H2S

ACK

Zero gas setting acknowledged

H2S

CXXX PPM

Banner: Apply span gas, then enter <E> to calibrate gas sensor

Sub-Routines

XXXPPM

Default Display

Operation: Confirm area clear of gas, or apply zero air to
sensor.

H2S

10PPM-SP

Table 6 - 1: Calibration