NI 5412 Specifications

Maximum Output Voltage

Maximum
Output
Voltage

Path

Load

Maximum Output Voltage (V

pk-pk

)

The Maximum
Output Voltage
of the NI 5412 is
determined by
the Amplitude
Range and the
Offset Range.

Low-

Gain

Amplifier

50

±1.000

1 k

±1.905

Open

±2.000

High-

Gain

Amplifier

50

±6.000

1 k

±11.43

Open

±12.00

Accuracy

DC Accuracy

±0.2% of Amplitude ± 0.05% of Offset ± 500 µV
(within ±10 °C of self-calibration temperature)

±0.4% of Amplitude ± 0.05% of Offset ± 1 mV
(0 °C to 55 °C)

Calibrated for
high impedance
load.

AC Amplitude
Accuracy

±1.0% of desired Amplitude ± 1 mV

50 kHz sine
wave.

Output Characteristics

Output
Impedance

50

Ω or 75 Ω nominal, software-selectable.

Output
Coupling

DC

Output Enable

Software-selectable. When the Output Path is disabled, the
CH 0 Output is terminated to ground with a 1 W resistor
equal to the selected output impedance.

Maximum
Output
Overload

The CH 0 output can be connected to a 50

Ω, ±12 V source

without sustaining any damage. No damage occurs if the
CH 0 output is shorted to ground indefinitely.

Waveform
Summing

The CH 0 output supports waveform summing among
similar paths—specifically, the outputs of multiple NI 5412
signal generators can be connected directly together.

Table 1. (Continued)