National Instruments NI 5412 User Manual
Page 8

NI 5412 Specifications
8
ni.com
Sample Rate Range and Resolution
Sample Clock
Source
Sample Rate Range
Sample Rate Resolution
—
Divide-by-N
23.84 S/s to 100 MS/s
Settable to (100 MS/s) / N
(1
≤ N ≤ 4,194,304)
High-
Resolution
10 S/s to 100 MS/s
1.06 µHz
CLK IN
200 kS/s to 105 MS/s
Resolution determined by
external clock source.
External Sample Clock duty
cycle tolerance 40% to 60%.
NI PXI-5412
PXI Star
Trigger
10 S/s to 105 MS/s
NI PXI-5412
PXI_Trig<0..7>
10 S/s to 20 MS/s
NI PCI-5412
RTSI<0..7>
10 S/s to 20 MS/s
Effective Sample Rate
Sample Rate
(MS/s)
Interpolation
Factor
Effective Sample
Rate
Effective Sample
Rate =
(Interpolation
Factor) * (Sample
Rate)
10 S/s to
105 MS/s
1 (Off)
10 S/s to
105 MS/s
12.5 MS/s to
105 MS/s
2
25 MS/s to
210 MS/s
10 MS/s to
100 MS/s
4
40 MS/s to
400 MS/s
10 MS/s to
50 MS/s
8
80 MS/s to
400 MS/s
Sample Clock Delay Range and Resolution
Sample Clock
Source
Delay Adjustment Range
Delay Adjustment
Resolution
—
Divide-by-N
±1 sample clock period
<10 ps
High-
Resolution
±1 sample clock period
Sample Clock
Period/16,384
Table 2. (Continued)
Specification
Value
Comments