Power, Input connector pin assignment – Omega Vehicle Security wall mount Multi-Channel DPS3300 User Manual

Page 25

Advertising
background image

Page 24

8

9

10

11

12

13

6

5

4

3

2

1

7

14


POWER


Power connection should be made to the three terminal connector as shown in figure 3. Also, make note that it is very
important that the power line inputs and the power ground are not switched. Doing so will permanently damage the
instrument. Refer to Figure 3 for proper connections. For convenience, the printed circuit board is labeled
L1 L2 GND on the under side of three terminal power connector. For instruments with the 15Vdc power option,
L1 = negative supply and L2 = positive supply.

NOTE: WHILE CONNECTING POWER TO THE UNIT, MAKE SURE THAT AC POWER LINE L1 OR
L2 IS NOT ACCIDENTALLY CONNECTED TO GROUND (GND) TERMINAL. THIS WILL RESULT IN
PERMANENT DAMAGE TO THE INSTRUMENT.

DOUBLE CHECK THE CONNECTIONS BEFORE APPLYING POWER!!


INPUT CONNECTOR PIN ASSIGNMENT





CH1

-

CH1+ CH2

-

CH2 + CH3

-

CH3 + CH4

-

CH4 + CH5

-

CH5 + CH6

-

CH6 + CH7

-

CH7 +





Table 4. DPS3300 Series Output Connector Pin Assignments

OUT-PUT CONNECTOR PIN ASSIGNMENTS

Pin # Signal

Relay Output

Open Collector

1 Channel #1 Limit

Common

Positive Output

2

Channel #1 Limit

Normally Open/Closed*

Negative Output

3

Channel #2 Limit

Common

Positive Output

4

Channel #2 Limit

Normally Open /Closed*

Negative Output

5

Channel #3 Limit

Common

Positive Output

6

Channel #3 Limit

Normally Open/Closed*

Negative Output

7

Channel #4 Limit

Common

Positive Output

8

Channel #4Limit

Normally Open/Closed*

Negative Output

9

Channel #5 Limit

Common

Positive Output

10 Channel #5 Limit

Normally Open/Closed*

Negative Output

11

Channel #6 Limit

Common

Positive Output

12 Channel #6 Limit

Normally Open/Closed*

Negative Output

13

Channel #7 Limit

Common

Positive Output

14

Channel #7 Limit

Normally Open/Closed*

Negative Output

* = Normally Open/Closed is user programmable (Look under Set-up)

NOTE: PROPER RATING, CONNECTION AND CORRECT ORIENTATION OF THE CONNECTOR IS
NECESSARY TO AVOID MALFUNCTION OR PERMANENT DAMAGE TO THE UNIT.