TROUBLE SHOOTING


If the meter locks on a particular display, unplug and plug in again to unlock the display. A total reset clears the memory
and defaults all values to zero. To perform this reset, re-apply power with RESET key pressed in.

Factory Channel settings:

To obtain factory preset values simply re-apply power with SDIGT key pressed in. This will also clear memory of any
erroneous values and is the preferred method of resetting a unit that might be malfunctioning due to corrupted memory .

Factory default settings for Thermocouple, voltage, current and milli-volt are:

CURRENT

VOLTAGE.

MILLIVOLT

T/C

PARAMETER

INPUT

INPUT INPUT

INPUT


HIGH SCALE

20.000

10.000

100.00

J

LOW SCALE

0.000

0.000

0.000

J

OFFSET

0.00

0.00

0.00

J

TARE

0.000

0.000

0.000

J


ERROR MESSAGES AND SOLUTIONS

ERROR MESSAGES AND SOLUTIONS


ERROR MESSAGE

ERROR CONDITION

SOLUTION


'EE EROR'

The unit was unable to

If the message keeps coming up,

save

programmed

value

unplug

the

unit

and

reapply

power.

in EEPROM -- possible

If the unit still displays the error

failure

of

EEPROM.

message,

send

it

in

for

repair.

If error message followed by
8.8.8.8.8.8.8.and the revision
number are displayed occasionally,
it is an indication that power

line

is

very

noisy

and

needs

a

line

filter.


'CARD ER'

Too much noise on

1. If message shows occasionally,

input signals or problem

it indicates that input signal s have

with

Scanner

card. too

much

high

voltage

noise.

Provide

signal

filters.

2.

Unplug

and

plug

the

meter

again

to recycle power.

3.

Scanner

card

defective

--

send

in

for repair


'HIGH AC NOISE'

High AC noise on input

1. If message shows occasionally,

signals or incoming AC it indicates that input signals or

line interfering with the AC line has too much high

operation of the unit

voltage noise.

2. Provide line and signal filters.