Particle Specifications
A series of aerosol performance tests were performed using an 8 stage cascade impactor at a
sampling flow rate of 15 l/min equipped with a USP <601> induction port throat. Aerosol was
sampled directly from the outlet. Three (3) device samples were tested with 3 runs each, for a total
of 9 sample points per each drug.

The specifications are listed below with intervals given for a 95% confidence level.

Mean / Std. Dev

Pulmicort

®

(250μg/mL)

Intal

®

(10mg/mL)

Salbutamol

®

(5mg/mL)

Total Delivered Dose (μg)

391.11 ± 16.51

12368.89 ± 269.61 7883.33 ± 116.96

Total Delivered Dose Fraction (%)

78.2% ± 3.3%

61.8% ± 3.1%

77.0% ± 2.5%

Particle size (MMAD) μm

3.88 ± 0.28

2.91 ± 0.14

2.54 ± 0.28

Geometric Standard Deviation

1.93 ± 0.20

2.27 ± 0.02

2.62 ± 0.03

Respirable Fraction (0.5-5μm)

61.9% ± 4.0%

70.5% ± 1.3%

66.4% ± 1.6%

Total Respirable Dose

(μg between 0.5-5μm)

242.17 ± 18.90

8729.49 ± 497.57

5235.02 ± 233.61

Medication Captured on USP Throat

13.34 ± 3.40

253.66 ± 27.42

144.94 ± 16.32

Medication Captured on USP

Throat Fraction (%)

3.4% ± 0.4%

2.0% ± 0.1%

1.8% ± 0.1%

Medication Retained in Device

96.11 ± 9.61

8000.00 ± 438.21

2072.22 ± 257.90

Medication Retained in Device

Fraction (%)

19.2% ± 1.9%

40% ± 2.2%

41.4% ± 5.2%

Coarse Particle Fraction (%) (>4.7μm)

37.1% ± 2.9%

27.7% ± 1.5%

26.6% ± 2.3%

Fine Particle Fraction (%) (<4.7μm)

59.6% ± 3.5%

70.2% ± 1.5%

71.6% ± 2.3%

Ultra-Fine Particle Fraction (%)

(<1.0μm)

3.5% ± 1.7%

8.9% ± 2.1%

16.9% ± 5.3%

Note: Course particles (oro-pharyngeal deposition) and ultra-fine particles (exhaled) are not likely

to deposit in the patient’s airway and thus provide limited clinical benefit.