Guidance and manufacturer's declaration – Omron NE-C801 User Manual

Page 30

Advertising
background image

30

Recommended separation distance between portable and mobile RF communications

equipment and the NE-C801
The NE-C801 is intended for use in an electromagnetic environment in which radiated RF

disturbances are controlled. The customers or the users of the NE-C801 can help prevent

electromagnetic interference by maintaining a minimum distance between portable and mobile

RF communications equipment (transmitters) and the NE-C801 as recommended below,

according to the maximum output power of the communications equipment.

Output Power of

Transmitter in Watt

Separation distance according to frequency of transmitter in meter
150 kHz to 80 MHz

d = 1.2 √P

80 kHz to 800 MHz

d = 1.2 √P

800 MHz to 2.5GHz

d = 2.3 √P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended

separation distance d in meters (m) can be estimated using the equation applicable to the

frequency of the transmitter, where P is the maximum output power rating of the transmitter

in watts (W) according to the transmitter manufacturer.

Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies

Note: These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects, and people.

EMC tests have included the AC adapter as included with the product.

GUIdANCE ANd MANUfACTURER'S dECLARATION