Omega TX82B User Manual

Page 13

Advertising
background image

9

6.0 CUSTOMER CONFIGURATION AND CALIBRATION

Use this procedure to determine the configuration of the TX82B customized
setup. The procedure is general; customers can specify any two current inputs
and their corresponding digital readings. Pin-groups are shown in Figure 5-1.

6.1 FORMULA

Base all your calculations on either the 1-5, 4-20 or 10-50 mA range.
1.

Determine the lowest input current, I1, which is specified by the customer:

I1 =

mA

2.

Determine the highest input current, I2, which is specified by the customer:

I2 =

mA

3.

Determine the reading, N1, at input current I1, which is specified by the
customer.

N1 =

counts

4.

Determine the reading, N2, at input current I2, which is specified by the
customer:

N2 =

counts

5.

Calculate the Gain, G1:

(N2 - N1)

G1 =

=

counts per mA

(I2 - I1)

(If G1 is greater than 125, it is out of range for a standard unit.)

6.

Calculate the Required Zero Range number, RZR:

RZR = N1 - (I1 x G1) =

7.

Select the Zero Range required from the following chart where the Required
Zero Range number falls between the upper and lower limits of that range:

ZR1 = -2510 to -1420
ZR2 = -1580 to -420
ZR3 = -470 to +850
ZR4 = +760 to +2000
ZR =

Advertising