2

2.0 SPECIFICATIONS

2.1 INPUT

Current

1-5 mA, 4-20 mA or 10-50 mA

Protection

200 mA max forward and 1000 mA max
reverse

Voltage Drop

2.5 V max forward and 1.2 V max reverse

Span Range

100 to 2000 counts continuous adjustment
with a multiturn potentiometer

Zero Range

The multiturn zero potentiometer can displace
the displayed reading by ±500 counts from the
center of each of the four overlapping zero
ranges, provided that the resultant is within the
±1999 count display capability.

Span Slope

Positive or Negative

Normal Mode Rejection

at 50/60 Hz

46 dB minimum

2.2 ACCURACY AT 25°C

Maximum error

±0.1% R ±1 count

Zero tempco

±0.1 ct/K typ, ±0.3 ct/K max

Span tempco

±0.005%S/K typ, ±0.015%S/K max

2.3 CONVERSION

Technique

Dual-slope, average value with autozero
correction

Polarity

Determined automatically at the end of input
integration period

Input integration period

100 milliseconds (nominal value)

Reading rate

2.5/second