5-6

1-5V Input

Select 1-5V.

Set the channel range.

Set the left span value.

Set the right span value.

Set the left scale value.

Set the right scale value.

Turn On or Off the low-cut function.

The new setting takes effect.

First channel

(Select the right scale value.)

Last channel

ESC/?

Set=Range

CH=01-01

Mode=1-5V

Span_L= 1,000

Span_R= 5,000

Scale_L= 0.00

Scale_R= 200.00

Low-cut=Off

01-01 Channel
Setting complete

Description

• Span Left and Span Right

Selectable span left range: 0.8 to 1.2 V

Selectable span right range: 4.8 to 5.2 V

• Scale Left and Scale Right

Selectable range (mantissa): –20000 to 30000

Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX

The decimal position is the position specified by scale left.

Make sure that scale left is less than scale right.

• Low-cut

On: Cuts values below 0% of the recording span (set to 0% (scale left value)).

Off: Does not apply the low-cut function.

Note

•

The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.

•

If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when the

low-cut function is applied is displayed as “–Over.”