Operation errors, Communication errors – Omega RD100B User Manual

Page 188

Advertising
background image

10-2

Code

Message

Explanation/Countermeasures

72

MATH channel is turned off

–

81

All space or ‘quit’ string cannot be specified.

–

86

The key-lock release password is incorrect.

Enter the correct password.

87

This key is locked.

–

91

Password is incorrect.

Enter the correct password.

100

IP address doesn’t belong to class A, B, or C.

–

101

The result of the masked IP address is all 0s or 1s.

–

102

SUBNET mask is incorrect.

Set a correct subnet mask.

103

The net part of default gateway is not equal to

Set the correct default gateway.

that of IP address.

151

This action is invalid during calculation.

–

160

This action is invalid during chart end.

–

161

This action is invalid during pen hold.

–

162

Cannot set an number for a skipped data.

–

163

This action is invalid during record.

–

164

This action is invalid during manual printing.

–

165

This action is invalid during list printing.

–

166

This action is invalid during setup list printing.

–

167

This action is invalid during chart feed.

–

169

This action is invalid during ribbon hold.

–

Operation Errors

Code

Message

Explanation/Countermeasures

232

There is no available data.

There is no data for periodic printout or data for calculating TLOG
when the timer expired.

Communication Errors

Code

Message

Explanation/Countermeasures

300

Command is too long.

–

301

Too many number of commands delimited with ’;’.

Keep the number of commands separated by sub delimiters
under 10.

302

This command has not been defined.

–

303

Data request command can not be enumerated with –
sub-delimiter.

350

Command is not permitted to the current user level.

–

351

This command cannot be specified in the current

–

mode.

352

The option is not installed.

–

353

This command cannot be specified in the current

–

setting.

354

This command is not available during calculation.

–

390

Command error.

–

391

Delimiter error.

–

392

Parameter error.

–

393

No permission.

–

394

No such connection.

–

395

Use “quit” to close this connection.

Attempted to disconnect its own connection.

396

Failed to disconnect.

–

397

No TCP control block.

The control block of the specified connection cannot be found.

10.1 A List of Error Messages

Advertising