Calibrating, Equipment needed for calibration – Quantum Data 700A User Manual

Page 9

Advertising
background image

9

Calibrating

Model 700A and 701A generators use programmable hardware to set the amplitude calibration

factors for analog video outputs. These calibration factors are stored in Flash EPROM, and are not

lost if the batteries are run down or removed.

Quantum Data calibrates the analog video and sync output amplitudes in new generators to pub-

lished specifications. You can recalibrate these amplitudes over a nominal range to meet your

specific testing conditions.

WARNING: Recalling the original factory calibration settings is not possible after saving

your own settings.

Equipment Needed for Calibration

DC voltmeter capable of measuring from 0 to 1 volt with a minimum resolution of 2 millivolts

Precision 75 ohm (+/- 1%) input terminator for the voltmeter

Test cable to connect the voltmeter to the R, G, B and ground pins on the VGA connector

Oscilloscope with a precision 75 ohm (+/- 1%) input impedance and suitable gain and band-

width to display a 1.00 Volt pk-pk signal with 3.579 MHz color carrier

75 ohm coaxial cable to connect the oscilloscope to the BNC TV output

Vector scope