Measuring Waveforms

3- 98

CSA8000B & TDS8000B User Manual

Performing a dark-level compensation maximizes the accuracy of the extinction
ratio and other optical automatic measurements you take. Performing a User
Wavelength Gain compensation optimizes an optical channel for your custom
input signal.

NOTE. Dark level compensation performs a subset of the module compensation
process. It is designed to be fast so it can be performed frequently, just before
measurements are taken. This compensation is not saved and are only valid for
the selected bandwidth or filter path and the internal optical power meter.

You should perform the procedure To Compensate the Instrument and Modules
on page 3--92 to compensate all vertical bandwidth and filter selections.

Use the following procedure to perform either compensation; this procedure
applies only to optical modules.

Overview

To perform optical compensations

Control elements and resources

Prerequisites 1.

The instrument must be installed with at least one
optical sampling modules to be dark-level calibrated in
place. The acquisition system should be set to run
continuously.

See the sampling-module User Manuals for

sampling module installation.

Select the

waveform

2.

Use the Vertical buttons to select the channel to be
compensated.

Access

dark-level com-

pensation

3.

From the application menu bar, click Setup, and then
click Vertical. See right.

To Perform Dark-Level

and User Wavelength Gain

Compensations