Trimble ThunderBolt E GPS Disciplined Clock User Guide

67

Trimble Standard Interface Protocol

A

Command Packet 0x8E-A0: Set DAC Value

Use command packet 0x8E-A0 to set the DAC output voltage or to request the current
DAC output voltage plus the parameters describing the DAC. The DAC output voltage
is used to control the frequency of the OCXO (Oven Controlled Crystal Oscillator.)
Send this packet with no data to request the DAC voltage. The receiver responds with
packet 0x8F-A0.

The Command Packet Data Fields are as follows:

To set the DAC voltage, send packet 0x8E-A0 in the format shown below.

Note – Oscillator disciplining must be disabled (see packet 0x8E-A3) to use this command
to set the DAC voltage.

Command Packet 0x8E-A0 Data Format

Command Packet 0x8E-A2: UTC/GPS Timing

Command packet 0x8E-A2 sets the UTC/GPS timing mode (time and date fields) in
packet 0x8F-AB, and the temporal location of ThunderBolt E output PPS. Send packet
0x8E-A2 with no data to request the current settings. The receiver replies with
response packet 0x8F- A2.

Field

Description

Setting

Voltage/Value flag

Use this field to specify that the DAC is to be
set either by value or by voltage.

0: Set DAC by voltage

1: Set DAC by value

DAC Voltage Value:

When the Voltage/Value Flag is set to
voltage, use this field to specify the numeric
value of the DAC as the 32-bit unsigned
number.

Voltage: 0.0 V to +4.0 V

Value: 0x00000 to
0xFFFFF (20 bits)

Byte

Item

Type

Value

Description

0

Subcode

UINT8

0xA0

1

Voltage /Value Flag

UINT8

0

1

Set DAC voltage

Set DAC value

2-5

DAC Voltage/ Value Single/UINT3 2

DAC voltage/Value

Byte

Bit

Item

Type

Value

Meaning

0

Subcode

UINT8

0xA2

1

0

1

UTC/GPS
time

bit field

0

1

GPS time/date in packet 0x8F-AB

UTC time/date in packet 0x8F-AB

bit field

0

1

PPS referenced to GPS time

PPS referenced to UTC time