Model 5100-02-IT Combustible Gas Sensor Module

Page: 40

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns contents of selected Modbus Register

Command 144: Set Transducer Serial Number

This command allows user to add a device specific serial number, if required. This is different from the serial
number assigned by the Manufacturer.

Request Data Bytes:

Byte

Format

Description

0-2

Unsigned-24 Serial

Number

Command Specific Response Data Bytes:

Byte

Format

Description

0-2

Unsigned-16

Returns new Serial Number

Command 145: Set Calibration Mode output current

Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Current output required during calibration, in
micro-amperes

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns new Calibration current value

Command 146: Set Trouble Mode output current

Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Current output required to indicate Trouble, in
micro-amperes

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns new Trouble current value

Command 149: Set Gas Factor

Request Data Bytes

Byte

Format

Description

0

Unsigned-8

Sets the Gas Factor value, range 30 to 200