6. Squeeze the bottle until a tiny drop

forms at the tip. Touch the drop to the
front edge of the yellow window of the
test strip. When you see flash, you
have enough control solution in the test
strip. Wipe the tip of the bottle with a
tissue then cap the bottle tightly.

7. The result appears on the display, along

with the control bottle symbol and
flashing “L.” Do not remove the test
strip yet. Press

once to mark the

result as a Level 1. If you tested the
Level 2 control, press

a second

time.