Calibration – American Magnetics 286 Multi-Sensor Liquid Level Instrument (CE-Marked) User Manual
Page 34

20
Calibration
Calibration Methods
Figure 3-2. Calibration method selection diagram.
Instrument &
sensor purchased
together?
START
Verify sensor
specifications
Verify calibration
sticker on bottom of
instrument
Perform
Open
Dewar Calibration
Is factory
calibration
correct?
Can
Open
Dewar Calibration be
performed in target
liquid?
Can
Closed
Dewar Calibration be
performed in target
liquid?
Perform
Closed
Dewar Calibration
Perform
Approximate
Calibration
Go to
Operation
Y
N
N
Y
N
Is factory
calibration
Approximate?
Y
Y**
N
** Factory Approximate Calibration should only be used if
open dewar or closed dewar calibration is not feasible in
the target liquid at the customer's facility.
N
Can
Open
Dewar Calibration be
performed in target
liquid?
Can
Closed
Dewar Calibration be
performed in target
liquid?
Y
Y
N
N
Y