Calibration
Calibration Menu Reference
and it must be re-entered. The capacitance of the MAX point must be
greater than the MIN point in order to perform level measurements.
The MIN point corresponds to the 0% level and minimum
capacitance. Press
7
to initiate the MIN measurement. The
instrument will emit a short beep if the attempt is successful and
the display will change to
Entered
.
The MAX point corresponds to the 100% level and maximum
capacitance. Press
8
to initiate the MAX measurement. The
instrument will emit a short beep if the attempt is successful and
the display will change to
Entered
.
Press the
ESC
key to exit the perform sub-menu once the MIN and
MAX measurements have been entered. Proceed to the Operations
section to configure the remaining instrument functions.